SF8368-B輸液器泄漏負壓測試儀
SF8368-BInfusion sets leakage negative pressure tester
SF8368-B輸液器泄漏負壓測試儀是依照GB8368-2005《性使用輸液器、重力輸液式》中的有關(guān)條款而設計制造的,用于檢測在輸液器內(nèi)部施加低于大氣壓強20kPa的氣壓,檢測有無氣體進入。
SF8368-B Infusion sets leakage negative pressure tester is designed and manufactured according to the relevant clauses of GB8368-2005《Infusion sets for single use, gravity feed》. It is mainly used for testing if there is leaked gas when infusion sets interior is born air pressure that lower 20kPa than atmosphere.
主要技術(shù)參數(shù):
Main technical parameters:
負壓輸出范圍:-10kPa~-30kPa(可調(diào)),LED數(shù)字顯示,誤差不大于&plun;0.5kPa。
Negative pressure output scope: -10kPa~-30kPa(adjustable), LED dial display, error≤5kPa.
測量系統(tǒng)負壓保持:在1分鐘內(nèi)不得飄移&plun;1kPa。
Test system negative pressure keep: It should not exceed&plun;1kPa in one minute.
測試持續(xù)時間:1S~60S(可調(diào)),LED數(shù)字顯示,誤差不大于&plun;1S。
Test keep time: 1S~60S(adjustable), LED dial display, error≤1S.