- 品牌/商標(biāo):梅特勒-托利多
- 原產(chǎn)地:瑞士
快速獲得稱量結(jié)果 – 價(jià)值永恒
無(wú)論是簡(jiǎn)單的日常工作,還是復(fù)雜的稱量程序,NewClassic MS半微量天平將卓越的性能和提高生產(chǎn)力進(jìn)行了完美的結(jié)合,這一切都來(lái)源于快速的穩(wěn)定時(shí)間以及一貫的可靠結(jié)果。來(lái)自瑞士的高分辨率稱量技術(shù)(HRT),保證了整個(gè)稱量范圍內(nèi)的始終。創(chuàng)新左右手互換開關(guān)門(ErgoDoor)使樣品的稱量盡可能簡(jiǎn)單和方便。內(nèi)置日期和時(shí)間功能,進(jìn)行符合GLP/GMP規(guī)范的打印輸出,確保您的研究的全面控制和可追溯性
Features and Benefits
Precise Results
The METTLER TOLEDO HRT (HiRes Technology) includes two built-in weights. Together with FACT (Fully Automatic Calibration Technology), this allows the balance to automatically perform adjustment and linearization.
Efficient Handling
Reduce the number of movements needed for every weighing. Open the draft shield ErgoDoor with one hand and simultaneously place the weighing sample with your other hand.
Space Saving
Save working space thanks to split draft shield doors that do not project beyond the balance, and sideways mounted connectors at the back of the balance.
Easy Cleaning
Cleaning has never been this easy: dismantle the QuickLock draft shield in a few short steps, then wash all glass panels and the drip tray in the dishwasher.
HCD (High Contrast Display)
FACT fully automatic time and temperature-controlled internal adjustment
Protective cover (terminal)
Protective cover
Below weighing hook for weighing below the balance
HRT weighing cell (HiRes Technology)
Overload Protection
Built in Date and Time function
SmartTrac shows the entire weighing range
Programmable 3 SmartKeys for shortcut access to applications
Protection of main balance configuration
Weighing in 16 different units
Taring over the entire weighing range
Be adapted to the environment
Applications like Piece counting, Percent weighing, Free factor, Statistics, Formulation, Totaling
Working in different languages: German, English, French, Spanish, Italian, Polish, Czech
Limit values
Maximum capacity 220 g
Maximum capacity, fine range 82 g
Readability 0.1 mg
Readability, fine range 0.01 mg
Repeatability (at nominal load) 1) 0.08 mg (200 g)
Repeatability (at low load) 1)
Linearity deviation 0.2 mg
Sensitivity offset 2) 0.8 mg (200 g)
Typical values
Repeatability (at nominal load) 1) 0.07 mg (200 g)
Repeatability (at low load) 1) 0.015 mg (20g)
Minimum weight (according to USP) 45 mg
Minimum weight (@U=1%, 2 sd) 3 mg
Settling time 4s / 8s
Sensitivity offset 2) 0.6 mg (200g)
Dimensions
Weighing pan dimensions (mm) ? 80
Balance dimensions LxWxH (mm) 358x247x331
1) Standard deviation (sd)
2) Stability of sensitivity with FACT self-adjustment switched on